IPEN Adolescent Accelerometer Data Collection Guidelines
Filter:
MUST USE the Low Frequency Extension with new generation Actigraphs (GT1M,
ActiTrainer, GT3X, GT3X+). Doesn’t apply to old generation devices.
Apply during initialization with GT1M & GT3X
Apply during or post-download with GT3X+
Kelli L. Cain, San Diego May 2014
The Coordinating Center conducted a
model/filter comparison study
27 adolescents wore a 7164 and GT3X+ on the
same belt for 3 days in free-living environment
Data are showing that the 7164 and
GT3X+Normal filter are not comparable (e.g., 9
min/day less MVPA with the GT3X+)
Applying the Low Frequency Extension reduces
these differences!
See LFE paper in adults:
Cain KL, Conway TL, Adams MA, Husak LE, Sallis JF.
Comparison of older and newer generations of
ActiGraph accelerometers with the normal filter
and the low frequency extension. International
Journal of Behavioral Nutrition and Physical
Activity 2013, 10:51.